Selective atrial sensing in dual chamber pacemakers eliminates endless loop tachycardia

Autor: Klementowicz, Peter T., Furman, Seymour
Zdroj: Journal of the American College of Cardiology; March 1986, Vol. 7 Issue: 3 p590-594, 5p
Abstrakt: With the introduction of dual chamber pacemakers that have multiple atrial amplitude sensing values, selective P wave sensing is possible. Five consecutive patients were studied who had 1) retrograde atrioventricular conduction, 2) anterograde atrial signals that were at least 1.4 times larger than their corresponding retrograde, atrial signals, and 3) dual chamber pulse generators that are capable of discriminating this difference in atrial amplitude. In each patient the pacemaker was programmed in the DDD mode and the posiventricular atrial refractory interval was at least 100- ms shorter than the individual's minimal retrograde conduction time. Two atria! sensitivity settings were evaluated in each patient: a high setting to ensure sensing of both anterograde and retrograde P waves, and a lower setting to allow sensing of anterograde P waves only. Ambulatory electrocardiographic monitoring demonstrated that with a high sensitivity setting, each patient sustained endless loop tachycardia (mean number of episodes 41, range 6 to 143) and that a logy atria! sensitivity setting eliminated the tachycardia. With the lower atrial sensitivity setting, there was only sporadic atrial undersensing (1.5 episodes for each 1,000 P waves).
Databáze: Supplemental Index