Positive correlation of optical density at 650 nm. with lecithin/sphingomyelin ratios in amniotic fluid

Autor: Sbarra, Anthony J., Selvaraj, Ratnam J., Cetrulo, Curtis L., Kennedy, Joseph L., Herschel, Marguerite J., Knuppel, Robert, Kappy, Kenneth, Mitchell, George W., Kelley, Edward C., Paul, Benoy B., Louis, Farid J.
Zdroj: American Journal of Obstetrics and Gynecology; April 1978, Vol. 130 Issue: 7 p788-790, 3p
Abstrakt: In this study, we have attempted to correlate optical density measurements of amniotic fluids with L/S ratios. We may conclude, with over a 98 per cent accuracy, that fluids having optical density readings of 0.15 and above, at 650 nm. will have an L/S ratio over 2.0. Fluids having optical density readings up to 0.05 will have L/S ratios of about 1.3. Finally, amniotic fluids having optical densities >0.05 and <0.15 will have L/S ratios of approximately 1.5.