Autor: |
Sbarra, Anthony J., Selvaraj, Ratnam J., Cetrulo, Curtis L., Kennedy, Joseph L., Herschel, Marguerite J., Knuppel, Robert, Kappy, Kenneth, Mitchell, George W., Kelley, Edward C., Paul, Benoy B., Louis, Farid J. |
Zdroj: |
American Journal of Obstetrics and Gynecology; April 1978, Vol. 130 Issue: 7 p788-790, 3p |
Abstrakt: |
In this study, we have attempted to correlate optical density measurements of amniotic fluids with L/S ratios. We may conclude, with over a 98 per cent accuracy, that fluids having optical density readings of 0.15 and above, at 650 nm. will have an L/S ratio over 2.0. Fluids having optical density readings up to 0.05 will have L/S ratios of about 1.3. Finally, amniotic fluids having optical densities >0.05 and <0.15 will have L/S ratios of approximately 1.5. |
Databáze: |
Supplemental Index |
Externí odkaz: |
|