Response characteristics of cellulose nitrate track-detectors to radon and radon-daughter mixtures

Autor: Knöfel, T.M.J., Estrada, J.J.S., Tavares, O.A.P., Martins, J.B., Goodwin, A.
Zdroj: Nuclear Instruments and Methods in Physics Research; July 1983, Vol. 212 Issue: 1-3 p387-392, 6p
Abstrakt: The response of bare cellulose nitrate (CN) LR-115, type II (Kodak-Pathé), track detector exposed to various mixtures of radon and airborne radon-daughters is investigated. Under controlled irradiation conditions good linear correlation was obtained between through-etched track density ϱ and the radon-daughter accumulated exposure ϵa(in units of WL-h). It is shown that the calibration factor Δϵa/Δϱfor the CN LR-115, type II, track detector is an increasing function of the equilibrium ratios of the radon-daughter mixtures.