Autor: |
Maguen, Ezra, Nesburn, Anthony B., Jackman, Norman, Macy, Jonathan I. |
Zdroj: |
American Journal of Ophthalmology; February 1985, Vol. 99 Issue: 2 p170-172, 3p |
Abstrakt: |
The haptic of a semiflexible anterior chamber intraocular lens (Kelman Quadraflex) broke inside the eye of a 70-year-old man. No history of major eye or head trauma could be elicited from the patient. Clinical manifestations included inferior corneal edema progressing toward the optical axis resulting from a free-floating haptic segment of the lens implant inside the anterior chamber. A significant decrease in endothelial density was documented over a short period of time. The surgical management of this case included removal of the free-floating haptic segment, removal of the lens implant, penetrating keratoplasty, and vitrectomy. Final visual acuity was 20/70. |
Databáze: |
Supplemental Index |
Externí odkaz: |
|