Autor: |
McPhate, Jason, Anne, Joshi, Bacinski, John, Banks, Bruce, Cates, Carey, Christensen, Paul, Cruden, Brett, Dunham, Larry, Graham, Eric, Hughes, David, Kimble, Randy, Lupie, Olivia, Niedner, Malcolm, Osterman, Steven, Penton, Steven, Proffitt, Charles, Pugel, Diane, Siegmund, Oswald, Wheeler, Thomas |
Zdroj: |
Proceedings of SPIE; September 2011, Vol. 8145 Issue: 1 p81450I-81450I-10, 8063561p |
Abstrakt: |
The color dependence of the measured decline of the on-orbit sensitivity of the FUV channel of the HST Cosmic Origins Spectrograph (HST-COS) indicated the principal loss mechanism to be degradation of the cesium iodide (CsI) photocathode of the open-faced FUV detector. A possible cause of this degradation is contamination by atomic oxygen (AO), prompting an investigation of the interaction of AO with CsI. To address this question, opaque CsI photocathodes were deposited on stainless steel substrates employing the same deposition techniques and parameters used for the photocathodes of the HST-COS FUV detector. The as-deposited FUV quantum efficiency of these photocathodes was measured in the 117-174 nm range. Several of the photocathodes were exposed to varying levels of thermalized, atomic oxygen (AO) fluence (produced via an RF plasma). The post AO exposure QE's were measured and the degradation of sensitivity versus wavelength and AO fluence are presented. |
Databáze: |
Supplemental Index |
Externí odkaz: |
|