Autor: |
Meiss, Thorsten, Rossner, Tim, Minamisava Faria, Carlos, Völlmeke, Stefan, Opitz, Thomas, Werthschützky, Roland |
Zdroj: |
Proceedings of SPIE; May 2011, Vol. 8068 Issue: 1 p806805-806805-13 |
Abstrakt: |
For the HapCath system, which provides haptic feedback of the forces acting on a guide wire's tip during vascular catheterization, very small piezoresistive force sensors of 200•200•640m3have been developed. This paper focuses on the characterization of the measurement performance and on possible new applications. Besides the determination of the dynamic measurement performance, special focus is put onto the results of the 3- component force vector calibration. This article addresses special advantageous characteristics of the sensor, but also the limits of applicability will be addressed. As for the special characteristics of the sensor, the second part of the article demonstrates new applications which can be opened up with the novel force sensor, like automatic navigation of medical or biological instruments without impacting surrounding tissue, surface roughness evaluation in biomedical systems, needle insertion with tactile or higher level feedback, or even building tactile hairs for artificial organisms. |
Databáze: |
Supplemental Index |
Externí odkaz: |
|