Autor: |
Bridwell, L., Maddox, W. E., Beyer, L. M., Etherton, R. C. |
Zdroj: |
Applied Optics; April 1970, Vol. 9 Issue: 4 p929-932, 4p |
Abstrakt: |
A McPherson 225 vacuum uv spectrometer has been interfaced with an IBM 1130 computer. The electronic interface, consisting mainly of integrated circuits, translates computer commands for a dc synchronous stepping motor which controls the grating position. Intensity information is recorded in digital form on a magnetic disk and may be plotted on a Calcomp plotter. A strip chart recorder is operated simultaneously. Reproducibility of the grating position as determined with a mercury source is within 0.03 Å on the average. |
Databáze: |
Supplemental Index |
Externí odkaz: |
|