Autor: |
Demirgian, Jack, Macha, Susan |
Zdroj: |
Field Analytical Chemistry & Technology; 1999, Vol. 3 Issue: 2 p95-104, 10p |
Abstrakt: |
Detection and quantitation of chemicals in the environment requires Fourier-transform infrared (FTIR) instruments that are properly calibrated and tested. This calibration and testing requires field testing using matrices that are representative of actual instrument use conditions. Three methods commonly used for developing calibration files and training sets in the field are a closed optical cell or chamber, a large-scale chemical release, and a small-scale chemical release. There is no best method. The advantages and limitations of each method should be considered in evaluating field results. Proper calibration characterizes the sensitivity of an instrument, its ability to detect a component in different matrices, and the quantitative accuracy and precision of the results. © 1999 John Wiley & Sons, Inc. Field Analyt Chem Technol 3: 95104, 1999 |