Is intrasound vibration useful in the diagnosis of occult scaphoid fractures?

Autor: Roolker, L., Tiel-van Buul, M.M.C., Broekhuizen, T.H.
Zdroj: Journal of Hand Surgery - American Edition; March 1998, Vol. 23 Issue: 2 p229-232, 4p
Abstrakt: This study was designed to confirm the results of Finkenberg et al. (J Hand Surg 1993; 18A: 4-7), who found a high sensitivity (100%) and specificity (95%) of the intrasound vibration method in diagnosing occult scaphoid fractures. These occult scaphoid fractures are not visible on x-ray films, but clinically the patients are suspected to having a scaphoid fracture. A vibratory apparatus is placed over the anatomical snuff-box and a vibration of 100 mW is emitted; a painful sensation is produced if the scaphoid is fractured. Thirty-seven consecutive patients with a clinically suspected scaphoid fracture were evaluated. In 6 patients, a scaphoid fracture was radiographically identified; in the remaining 31 patients, a 3-phase bone scan was obtained. Eleven wrists showed increased uptake over the scaphoid and were considered to have an occult scaphoid fracture. In this group, bone scintigraphy was used as the reference standard. The vibration test was painful in 1 of 6 patients with a proven scaphoid fracture and in 3 of the 11 patients with a positive bone scan. In contrast to the results of Finkenberg et al, the intrasound vibration method shows a sensitivity of 24%, a specificity of 85%, a positive predictive value of 40%, and a negative predictive value of 65%. We conclude that the accuracy of intrasound vibration is low and that it is not useful in the diagnosis of scaphoid fractures.
Databáze: Supplemental Index