A Real-Time PCR Method to Rapidly Titer Adenovirus Stocks.

Autor: Walker, John M., Tollefson, Ann E., Thomas, Maria A., Lichtenstein, Drew L., Krajcsi, Peter, Wold, William S. M.
Zdroj: Adenovirus Methods & Protocols; 2007, p185-192, 8p
Abstrakt: A critical step in working with adenovirus (Ad) and its vectors is the accurate, reproducible, sensitive, and rapid measurement of the amount of virus present in a stock. Titration methods fall into one of two categories: determination of either the infectious or the particle (infectious plus noninfectious) titer. Determining the infectious titer of a virus stock by plaque assay has important limitations, including cell line-, researcher-, and laboratory-dependent variation in titer, and the length of time required to perform the assay (2-4 wk). A major drawback of particle titration methods is the lack of consistent correlation between the resultant titer and the infectious titer. To overcome these problems, a rapid, sensitive, and reproducible real-time polymerase chain reaction (PCR) assay was developed that detects encapsidated full-length genomes. Importantly, there is a linear correlation between the titer determined by the real-time PCR assay and the infectious titer determined by a plaque assay. This chapter provides step-by-step guidance for preparing viral DNA, conducting the real-time PCR assay, and using the resultant data to calculate a viral titer. [ABSTRACT FROM AUTHOR]