Abstrakt: |
Purpose: To evaluate the reproducibility and reliability of a novel electromyogram (EMG) device with a flexible sheet sensor for measuring muscle activity related to mastication and swallowing. Methods: We developed a new EMG device made of elastic sheet electrodes to measure the masseter and digastric muscle activities for evaluating mastication and swallowing. To examine the measurement reproducibility of the new EMG device, masseter muscle activity was analyzed using the intraclass correlation coefficient (ICC). Further, we measured the maximum amplitude, duration, integrated value, and signal-to-noise ratio (SNR) using the new EMG device and conventional EMG devices and evaluated the reliability using ICC and Bland-Altman analysis. Results: We confirmed high ICC (1,1) and ICC (2,1) scores (0.92 and 0.88, respectively) while measuring the reproducibility of the new EMG device. When compared to the active electrode EMG device, we found a high correlation for the maximum amplitude (0.90), duration (0.99), integrated values (0.90), and SNR (0.75), with no observation of significant fixed errors. Moreover, the regression coefficient was not significant for any of the evaluation items and no proportional error was observed. Compared with the passive electrode EMG device, the maximum amplitude and duration were highly correlated (0.73 and 0.89). In addition, the SNR exhibited a significant fixed error. In contrast, the regression coefficient was not significant for any of the evaluation items and no proportional error was observed. Conclusions: Our results suggest that the new EMG device can be used to reliably and reproducibly evaluate muscle activity during mastication and swallowing. [ABSTRACT FROM AUTHOR] |