Autor: |
Lenis, Tamara L., Botsford, Benjamin E., Kovacs, Kyle D., Papakostas, Thanos D. |
Zdroj: |
Retinal Cases & Brief Reports; Jan2024, Vol. 18 Issue 1, p91-93, 3p |
Abstrakt: |
Purpose: To report a technique to remove a dislocated ganciclovir implant in the vitreous cavity. Methods: Retrospective case series. Two patients with dislocated ganciclovir implants in the vitreous cavity. Results: A 6-mm pars plana incision was made; the soft tip was used to elevate the implant behind the intraocular lens and then 0.12-mm forceps were used to externalize the implant. The implant was successfully removed in both patients. Conclusion: Removal of a dislocated ganciclovir implant with its encasing strut can be effectively retrieved using a bimanual approach. [ABSTRACT FROM AUTHOR] |