Abstrakt: |
Considering the importance to understand the behavior of the elements on different natural and/or anthropic processes, this study had as objective to verify the accuracy of a multielement analysis method for rocks characterization by using soil standards as calibration reference. An EDXRF equipment was used. The analyses were made on samples doped with known concentration of Mn, Zn, Rb, Sr and Zr, for the obtainment of the calibration curves, and on a certified rock sample to check the accuracy of the analytical curves. Then, a set of rock samples from Rio Bonito, located in Figueira city, Paraná State, Brazil, were analyzed. The concentration values obtained, in ppm, for Mn, Rb, Sr and Zr varied, respectively, from 175 to 1084, 7.4 to 268, 28 to 2247 and 15 to 761. [ABSTRACT FROM AUTHOR] |