Autor: |
Popov, A. N., Antson, A. A., Bondarenko, K. P., Belyaev, V. V., Harutunyan, E. G., Kheiker, D. M., Sheromov, M. A., Mytnytchenko, S. V. |
Předmět: |
|
Zdroj: |
Review of Scientific Instruments; Jul1989, Vol. 60 Issue 7, p2440, 1p |
Abstrakt: |
The station designed to speed up the collection of integral intensity data from protein and virus crystals is constructed on the SR beam of the VEPP-3 storage ring. A flat triangular Si crystal cut at 8° relative to the ( 111 ) plane is used to monochromatize the radiation. The monochromator is placed on the automatic goniometer which allows us to obtain the necessary energy of radiation, to make the adjusting turns and displacements, and to bend the crystal in order to achieve focusing in the horizontal plane. All the mechanical drive units are supplied with position sensors based on multiturn potentiometers. A conventional Arndt-Wonacott rotation camera is used to collect structure information. The station also includes an optical bench which can turn about the axis coinciding with that of the monochromator, and a set of slits to cut the background and to form the beam. The ionization camera serves to adjust the monochromator and to monitor the beam. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|