Autor: |
Silva, R. da, Veltcheva, R. I., Gray, J., Pearce, J. V., Machin, G., Teixeira, R. N. |
Předmět: |
|
Zdroj: |
AIP Conference Proceedings; Sep2013, Vol. 1552 Issue 1, p777-781, 5p, 1 Diagram, 6 Charts |
Abstrakt: |
In April 2011, a bilateral comparison of tin fixed point cells (231.928 °C) took place at the facilities of the Temperature group at the National Physical Laboratory (NPL). The comparison was with a standard from the Thermal Metrology Division of the National Institute of Metrology, Quality and Technology (INMETRO). It involved two cells: the test cell was an open one constructed in late 2010 at the INMETRO Thermometry Laboratory as part of an MSc research project; the reference cell was a sealed one, constructed in 2000 at NPL, traceable to the UK national standard tin point. The materials employed in the construction of the cells were from different suppliers. The cell design, dimensions and construction procedures were also different. Three standard platinum resistance thermometers (SPRTs), each different models from different manufacturers, were used to undertake the comparison, one from INMETRO and two from NPL. The comparison was performed in quadruplicate, each combination using a different freezing plateau. The differing self-heating behaviour of the different SPRTs was taken into account. The methodology employed in this comparison is detailed in the present paper. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|