Autor: |
Vautier, Julien, El Tannir El Tayara, Nadine, Walczak, Christine, Mispelter, Joël, Volk, Andreas |
Zdroj: |
Magnetic Resonance in Medicine; Sep2013, Vol. 70 Issue 3, p823-828, 6p |
Abstrakt: |
The purpose of this study was to provide proof of concept for a new three-dimensional (3D) radial dynamic contrast enhanced MRI acquisition technique, called 'Radial Entire Tumor with Individual Arterial input function dynamic contrast-enhanced MRI' (RETIA dynamic contrast-enhanced MRI), which allows for the simultaneous measurement of an arterial input function in the mouse heart at 2 s temporal resolution and coverage of the whole tumor. Alternating 2D and 3D projections contribute to the 2D heart image or 3D tumor data with a 3-cm field of view. Sixty-four 2D images of the heart are obtained during acquisition of each 3D tumor dataset. In a pilot study, global Ktrans and ve values were measured in four mice, in a respiratory motion-animated subcutaneously implanted breast tumor model. This technique is expected to be most useful for the characterization of microvasculature in motion-animated orthotopic tumors. Magn Reson Med 70:823-828, 2013. © 2012 Wiley Periodicals, Inc. [ABSTRACT FROM AUTHOR] |