Autor: |
Rodriguez Aguada, Sebastian, Rodriguez, Conrado J., Bruni, Rodrigo G., Vanella, Oscar R. |
Zdroj: |
IEEE Latin America Transactions; Feb2013, Vol. 11 Issue 1, p149-154, 6p |
Abstrakt: |
Although electrical medical equipments meet rigorous standards that ensure the safety and essential performance of the device, its needed to verify their status throughout serviceable lifespan. One of the parameters to check is the leakage current. The present work describes the development of an instrument for measuring total leakage current by differential method intended to verify the safety requirements established by international standard IEC62353. [ABSTRACT FROM PUBLISHER] |
Databáze: |
Complementary Index |
Externí odkaz: |
|