Volumetric imaging using single chip integrated CMUT-on-CMOS IVUS array.

Autor: Tekes, Coskun, Zahorian, Jaime, Gurun, Gokce, Satir, Sarp, Xu, Toby, Hochman, Michael, Degertekin, F. Levent
Zdroj: 2012 Annual International Conference of the IEEE Engineering in Medicine & Biology Society; 1/ 1/2012, p3195-3198, 4p
Abstrakt: An intravascular ultrasound (IVUS) catheter that can provide forward viewing volumetric ultrasound images would be an invaluable clinical tool for guiding interventions. Single chip integration of front-end electronics with capacitive micromachined ultrasonic transducers (CMUTs) is highly desirable to reduce the interconnection complexity and enable miniaturization in IVUS catheters. For this purpose we use the monolithic CMUT-on-CMOS integration where CMUTs are fabricated directly on top of pre-processed CMOS wafers. This minimizes parasitic capacitances associated with connection lines. We have recently implemented a system design including all the required electronics using 0.35-µm CMOS process integrated with a 1.4-mm diameter CMUT array. In this study, we present the experimental volumetric imaging results from an ex-vivo chicken heart phantom. The imaging results demonstrate that the single-chip forward looking IVUS (FL-IVUS) system with monolithically integrated electronics has potential to visualize the front view of coronary arteries. [ABSTRACT FROM PUBLISHER]
Databáze: Complementary Index