Autor: |
Gudino, N., Griswold, M. A. |
Zdroj: |
Magnetic Resonance in Medicine; Apr2013, Vol. 69 Issue 4, p1180-1185, 6p |
Abstrakt: |
Purpose A multi-turn transmit surface coil design was presented to improve B1 efficiency when used with current source amplification. Methods Three different coil designs driven by an on-coil current-mode class-D amplifier with current envelope feedback were tested on the benchtop and through imaging in a 1.5 T scanner. Case temperature of the power field-effect transistor at the amplifier output stage was measured to evaluate heat dissipation for the different current levels and coil configurations. In addition, a lower power rated device was tested to exploit the potential gain in B1 obtained with the multi-turn coil. Results As shown both on the benchtop and in a 1.5 T scanner, B1 was increased by almost 3-fold without increasing heat dissipation on the power device at the amplifier's output using a multi-turn surface coil. Similar gain was obtained when connecting a lower power rated field-effect transistor to the multi-turn coil. Conclusion In addition to reduce heat dissipation per B1 in the device, higher B1 per current efficiency allows the use of field-effect transistors with lower current ratings and lower port capacitances, which could improve the overall performance of the on-coil current source transmit system. Magn Reson Med 69:1180-1185, 2013. © 2013 Wiley Periodicals, Inc. [ABSTRACT FROM AUTHOR] |