Undersensing by an ICD Due to Alternans of the Ventricular Electrogram.

Autor: Van Heuverswyn, Frederic E., Timmers, Liesbeth, Stroobandt, Roland X., Serge Barold, S.
Zdroj: Annals of Noninvasive Electrocardiology; Jan2013, Vol. 18 Issue 1, p84-89, 6p
Abstrakt: Alternans of the ventricular electrogram (VEGM) during ventricular tachycardia (VT) is a rare cause of ventricular undersensing by an implantable cardioverter-defibrillator (ICD). This report describes a patient with a St. Jude ICD who exhibited sustained monomorphic VT associated with surface QRS alternans, alternating cycle lengths, alternans of the VEGM causing intermittent undersensing of the smaller component, and intermittent 2:1 counting of ventricular intervals during 1:1 sensing in response to the ICD detection algorithm. VEGM undersensing was corrected noninvasively simply by programming the threshold start from 62.5% to 50% which increased the sensitivity based on the amplitude of the VEGM. This maneuver did not affect the satisfactory and stable defibrillation threshold. [ABSTRACT FROM AUTHOR]