Measurement of internal left ventricular diameter by a tracking sonomicrometer.

Autor: STINSON, EDWARD B., RAHMOELLER, GLENN, TECKLENBERG, PAUL L.
Zdroj: Cardiovascular Research; Mar1974, Vol. 8 Issue 2, p283-289, 7p
Abstrakt: An ultrasonic system for the measurement of internal left ventricular transverse diameter in animals is described. The ventricular endocardial transducers are implanted easily and with minimal trauma via the left atrium. A tracking gate, instead of a conventional threshold detector, is used to follow the received signal. Reliable signals, including the first derivative of diameter, have been obtained in 90% of subjects. [ABSTRACT FROM PUBLISHER]