Autor: |
Morton Caldwell, W., Ebeling, Thomas, Patterson, R., Stradal, L. |
Zdroj: |
Medical & Biological Engineering (0025696X); 1973, Vol. 11 Issue 4, p500-505, 6p |
Abstrakt: |
A convertor to adapt general-purpose or obsolete oscilloscopes for biopotential measurements has been designed. This device consists of complete vertical and horizontal circuits to drive directly cathode-ray-tube delection plates. Vertical sensitivity ranges from 0·1 to 10 mV/cm and sweep speeds from 0·5 s/cm to 0·1 ms/cm. The linear-sweep sawtooth waveform is generated by an analogue integrator, which is reset by a vertical or external-input trigger signal. The circuits use low-cost operational amplifiers for design simplicity. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|