Progress in Electromagnetics Research Letters; 2012, Vol. 29, p115-126, 12p |
Abstrakt: |
Recently, we proposed a wireless ambulatory gait analysis system that provides a high ranging accuracy using ultra- wideband (UWB) transceivers. In this paper, we further investigate the performance of our proposed system including ranging using suboptimal templates, power consumption, and sensor-fusion. We show that the proposed system is capable of providing a 1.1mm ranging accuracy (11.7mm for current systems) at a signal-to-noise-ratio (SNR) of 20 dB using suboptimal-based receivers in industry accepted body-area-network UWB channels. For the angular-displacement, our system provides an accuracy that is less than 1° for the knee- flexion angle. This accuracy is superior to the accuracy reported in the literature for current technologies (less than 4°). Finally, we propose the integration of UWB sensors with force sensors. The system performance and design parameters are investigated using simulations and actual measurements. Ultimately, the proposed system is suitable for taking accurate measurements, and for tele-rehabilitation. [ABSTRACT FROM AUTHOR] |