Autor: |
Volodin, V., Souchkevitch, G., Racoveanu, N., Bergmann, H., Busemann-Sokole, E., Delaloye, B., Dermentzoglou, F., Georgescu, G., Herrera, N., Jasinski, W., Kasatkin, Y., Paras, P., Mould, R. |
Zdroj: |
European Journal of Nuclear Medicine; Mar1985, Vol. 10 Issue 5/6, p193-197, 5p |
Abstrakt: |
Twelve countries participated in the WHO inter-comparison for which transmission CAP (College of American Pathologists) brain and CAP liver phantoms and emission London liver phantoms were used. A total of 157 imaging devices were tested. Overall results from the phantoms revealed a wide range of targets detected. For the CAP-phantoms target detectability dropped below 50% for target sizes less than 9 mm and below 60% for target contrast less than 0.84:1. On average one false positive and six false negative results were reported using CAP-brain phantoms and one false positive and one false negative result using CAP-liver phantoms. For the London liver phantoms containing the tissue equivalent rubber abdominal simulation the target of 1 cm was never visualised. Two targets in this phantom (2 cm and 2.5 cm) were correctly identified in 34% of studied and one of these targets in 52% of studies. Equivocal and false positive results were reported in 42%. The WHO inter-comparison demonstrated the need to establish new, or to improve the existing, quality control programmes. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |