Lateral dose profile characterization in scanning particle therapy.

Autor: Bäumer, Christian, Farr, Jonathan B.
Předmět:
Zdroj: Medical Physics; Jun2011, Vol. 38 Issue 6, p2904-2913, 10p
Abstrakt: Purpose: In light-ion beam dose delivery with the scanning technique the spacing between adjacent spots is an important parameter during treatment planning. In order to study the effect of spot spacing on dose conformity and robustness for single field uniform dose configurations, fundamental geometrical properties of placement of Gaussian beamlets are explored. In particular, the dependence of penumbra width and flatness on spot width and spot spacing is investigated. Methods: Infinitesimal calculus and analytical methods are used to derive simple expressions for the lateral penumbra and the flatness of one-dimensional dose profiles in continuous scanning and uniform discrete spot scanning. In the same way expressions for the fundamental modes of perturbation of the spot sequence are developed. A numerical, matrix-based approach is followed to optimize weights spot-by-spot. Results: Generally the lateral penumbra widths lie between 1.13σb and 1.68σb with σb being the standard deviation of the beam spot profile. For regularly placed spots of equal weight with spot spacing λ the lateral penumbra is given by 1.68σ where σ results from quadratic subtraction of λ/12 from σb. The quantization error is identified as additional parameter describing the lateral dose conformity. It's variance is given by λ2/12 for a bunch of spots with uniform weights. The matrix-based optimization of weights for a one-dimensional dose box results in a lateral penumbra of typically 1.4σb. This value reduces to about 1.3σb if also the positions of the beam spots are optimized for the considered field size. The analytical formulas for uniform discrete scanning can be used as rough approximations of the best-case scenarios for weight-optimized dose profiles if the spot spacing is defined as effective spot spacing. Conclusions: The trade-off between flatness, quantization error, and robustness on the one side and penumbra width on the other side can be described analytically for equally weighted spots. Treatment planning systems often perform a least-squares optimization of the individual spot weights which results in smaller lateral penumbras and smaller quantization errors than for uniform discrete scanning. However, the benefit of this weight optimization decreases with increasing λ (in the regime λ>σb). The spot spacing, which is obtained from the scenario that the optimization objective is met by uniform discrete scanning, poses a sharp upper limit for the spot spacing λ in weight optimization methods. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index