Study of the uncertainty in NO chemiluminescence measurements due to the NO-O reaction in sampling lines.

Autor: Miñarro, Marta Doval, Ferradás, Enrique González, Rico, Jonathan Barberá, Alonso, Fernando Díaz, Martínez, Francisco J. Marzal, Trigueros, Cristina Romero
Předmět:
Zdroj: Environmental Science & Pollution Research; Mar2011, Vol. 18 Issue 3, p436-445, 10p
Abstrakt: Introduction: The change in light intensity that takes place when an ambient air sample is drawn into the detection chamber of a chemiluminescence monitor generates changes in the concentrations of several species, such as NO, NO and O. Although this phenomenon has been known for several decades, there is still no commonly accepted approach on when or how to correct for it in NO and O readings. Discussion: In this work, we have assessed the expanded uncertainty of two chemiluminescence NO analysers commercially available according to EN 14211:2005, with the aim of establishing the maximum allowable standard uncertainty due to the reaction between NO and O in the sampling system. Conclusion: Although this maximum allowable uncertainty cannot be a universal value-as it will depend on the performance of each analyser-our results have led us to propose the conservative value of 2%. We have also proposed a methodology for improving data quality which could be easily implemented by those responsible for air quality data validation. [ABSTRACT FROM AUTHOR]