Autor: |
Tamborini, Gabriele, Betti, Maria |
Zdroj: |
Microchimica Acta; Apr2000, Vol. 132 Issue 2-4, p411-417, 7p |
Abstrakt: |
Secondary ion mass spectrometry (SIMS) was optimised for characterisation of uranium- and plutonium-containing particles in soils, swipes and forensic samples. This was done by analysing in-house produced spherical UO2-particles. Screening techniques as α-autoradiography together with SIMS analysis were employed to detect UO2-particles in a soil sample from Chernobyl. The use of SIMS was exploited for the identification of uranium- and plutonium-containing particles and for the determination of their isotopic composition. The particles collected on swipe samples were transferred to a special adhesive support for the analysis by SIMS. Particles containing highly enriched uranium with diameters up to 10 μm were also detected in a forensic sample. For the measurements of the isotopic ratios a mass resolution of 1000 was used. At this resolution flat-top peaks were obtained which greatly improve the accuracy of the measurement. The isotopic composition of the particles was measured with a typical accuracy and precision of 0.5%. Statistically meaningful results can be obtained, for instance, from a specimen containing as few as 1010 atoms/μm3 of uranium in particles of UO2 weighing a few picograms. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|