Abstrakt: |
Two typical methods used for the determination of uranium in human autopsy tissues are kinetic phosphorescence analysis (KPA) and alpha-spectrometry, both of which have significant limitations and advantages. KPA is limited because of the amount of sample used (1–10 ml for sample digestion followed by one ml KPA aliquots), no isotopic information is provided, phosphorescence degradation by salts in solution, and even more importantly, it does not provide chemical recovery information. For samples with sub ng uranium concentrations per g of inorganic material, preconcentration is necessary, which may require chemical recovery (other than simple evaporation). While alpha-spectrometry has very good radiometric detection limits for 238U, the very long half-life of 238U (4.468·109 y) restricts its mass detection limit (27 ng). KPA, on the other hand, has a detection limit three orders of magnitude lower (0.02 ng) for natural uranium. A recovery corrected method for the determination of natural uranium in human tissues was developed combining preconcentration of human tissues dissolved in 6M HCl by anion exchange with alpha-spectrometry and kinetic phosphorescence analysis, utilizing 232U as a tracer. Solution aliquots containing up to 6 g of bone ash were pre-concentrated for KPA measurement thereby allowing the use of up to 25% of the original sample solution weight for analysis by KPA. The radiochemical yield of 232U was determined by alpha-spectrometry and the uranium content was determined by KPA. The mean radiochemical yields obtained for human tissue samples range from 65% to 106% with a mean of 85%±8%. [ABSTRACT FROM AUTHOR] |