A new validated high-performance liquid chromatographic method for the determination of EGIS-9933 in rat plasma.

Autor: Nemes, K., Szúnyog, J., Klebovich, I.
Zdroj: Chromatographia; Jul1998, Vol. 48 Issue 1/2, p158-162, 5p
Abstrakt: A new highly sensitive high-performance liquid chromatographic (HPLC) procedure for determination of EGIS-9933 (a newly developed anxiolytic compound) in rat plasma is described. A gradient, elution method with UV detection at 270 nm has been developed using a mobile phase of a mixture of A: methanol:acetonitrile 1:9 and B:0.5% triethilamine in water, the pH of B was adjusted to 3 with phosphoric acid. Solid phase extraction (SPE) was used for the sample preparation. The calibration was linear in the 10–10000 ng mL−1 concentration range. The limit of quantification was 10 ng mL−1. The bioanalytical method was validated according to internationally accepted criteria for biological samples. [ABSTRACT FROM AUTHOR]