Autor: |
Seely, John, Back, Christina, Deslattes, Richard, Hudson, Lawrence, Holland, Glenn, Bell, Perry, Miller, Michael |
Předmět: |
|
Zdroj: |
Review of Scientific Instruments; Jun2001, Vol. 72 Issue 6, p2562, 4p |
Abstrakt: |
A National Ignition Facility (NIF) core diagnostic instrument has been designed and will be fabricated to record x-ray spectra in the 1.1-20.1 keV energy range. The High-Energy Electronic X-Ray (HENEX) instrument has four reflection crystals with overlapping coverage of 1.1-10.9 keV and one transmission crystal covering 8.6-20.1 keV. The spectral resolving power varies from approximately 2000 at low energies to 300 at 20 keV. The spectrum produced by each crystal is recorded by a modified commercial dental x-ray charge coupled device detector with a dynamic range of at least 2500. © 2001 American Institute of Physics. [ABSTRACT FROM AUTHOR] |