Tibia fracture following removal of the ETN® (Expert Tibia Nail): a case report.

Autor: Seebauer, Christian J., van Scherpenzeel, Karine M., Haas, Norbert P., Bail, Hermann J.
Předmět:
Zdroj: Archives of Orthopaedic & Trauma Surgery; Jul2009, Vol. 129 Issue 7, p949-953, 5p, 1 Color Photograph, 5 Black and White Photographs, 3 Charts
Abstrakt: We report on a patient who sustained a fracture of the tibial shaft during the removal of the newest type of an intramedullary nail (Expert Tibia Nail, Synthes®). In this case report, we discuss the causes of this complication and possible ways to prevent this. [ABSTRACT FROM AUTHOR]