No Apparent Progress in Bioelectrical Impedance Accuracy: Validation Against Metabolic Risk and DXA.

Autor: Hemmingsson, Erik, Uddén, Joanna, Neovius, Martin
Předmět:
Zdroj: Obesity (19307381); Jan2009, Vol. 17 Issue 1, p183-187, 5p, 4 Charts, 2 Graphs
Abstrakt: Bioelectrical impedance (BIA) is quick, easy, and safe when quantifying fat and lean tissue. New BIA models (Tanita BC-418 MA, abbreviated BIA8) can perform segmental body composition analysis, e.g., estimate %trunkal fatness (%TF). It is not known, however, whether new BIA models can detect metabolic risk factors (MRFs) better than older models (Tanita TBF-300, abbreviated BIA4). We therefore tested the correlation between MRF and percentage whole-body fat (%BF) from BIA4 and BIA8 and compared these with the correlation between MRF and dual-energy X-ray absorptiometry (DXA, used as gold standard), BMI and waist circumference (WC). The sample consisted of 136 abdominally obese (WC ≥ 88 cm), middle-aged (30–60 years) women. MRF included fasting blood glucose and insulin; high-density lipoprotein cholesterol, low-density lipoprotein cholesterol, and triglycerides; high sensitive C-reactive protein, plasminogen activator inhibitor-1 (PAI-1), and fibrinogen; and alanine transaminase (ALT) liver enzyme. We found that similar to DXA, but in contrast to BMI, neither %BF BIA4 nor %BF BIA8 correlated with blood lipids or ALT. In the segmental analysis of %TF, BIA8 only correlated with inflammatory markers, but not insulin, blood lipids, or ALT liver enzyme (in contrast to WC and %TF DXA). %TF DXA was associated with homeostatic model assessment insulin resistance (HOMA-IR) independently of WC (P = 0.03), whereas %TF BIA8 was not (P = 0.53). Receiver-operating characteristic (ROC) curves confirmed that %TF BIA8 did not differ from chance in the detection of insulin resistance (P = 0.26). BIA estimates of fatness were, at best, weakly correlated with obesity-related risk factors in abdominally obese women, even the new eight-electrode model. Our data support the continued use of WC and BMI.Obesity (2008) 17 1, 183–187. doi:10.1038/oby.2008.474 [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index