Compact And Robust Laser Impulse Measurement Device, With Ultrashort Pulse Laser Ablation Results.

Autor: Kremeyer, Kevin, Lapeyre, John, Hamann, Steven
Předmět:
Zdroj: AIP Conference Proceedings; 4/28/2008, Vol. 997 Issue 1, p147-158, 12p, 1 Black and White Photograph, 1 Diagram, 1 Chart, 10 Graphs
Abstrakt: An impulse measurement device and analysis package was conceived, designed, constructed, tested, and demonstrated to be capable of: measuring nanoNewton-seconds to milliNewton-seconds of impulse due to laser-ablation; being transported as carry-on baggage; set-up and tear-down times of less than an hour; target exchange times of less than two minutes (targets can be ablated at multiple positions for thousands of shots); measurements in air and in vacuum; error of just a few percent; repeatability over a wide range of potential systematic error sources; and time between measurements, including ring-down and analysis, of less than 30 seconds. The instrument consists of a cantilever (i.e. leaf spring), whose time-dependent displacement/oscillation is measured and analyzed to determine the impulse imparted by a laser pulse to a target. These shapes are readily/commercially available, and any target material can be used, provided it can be fashioned in the form of a cantilever, or as a coating/film/tape, suitable for mounting on a cantilever of known geometry. The instrument was calibrated both statically and dynamically, and measurements were performed on brass, steel, and Aluminum, using laser pulses of ∼7 ns, ∼500 ps, and ∼500 fs. The results agree well with those published in the literature, with surface effects, atmosphere, and pre-/post-pulses demonstrating interesting effects and indicating areas for further study. These parameters should be carefully controlled and held constant during a series of measurements. The impulse imparted by ablation due to laser filaments in air was also explored. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index