Autor: |
Nina Tilly, Erik Grusell, Peter Kimstrand, Stefan Lorin, Konrad Gajewski, Mikael Pettersson, Andreas B, äcklund and, Bengt Glimelius |
Předmět: |
|
Zdroj: |
Physics in Medicine & Biology; May2007, Vol. 52 Issue 10, p2741-2754, 14p |
Abstrakt: |
In this paper we present the recent developments made for the scanning system for proton beams at TSL in Uppsala, showing that this system is now fully functional being able to produce conformal intensity modulated scan patterns with sufficient accuracy. A new control and supervising system handling the beam delivery including the control of the synchrocyclotron and the scanning system is developed and described in detail. A complete dosimetry system with transmission ionization chambers and a multi-wire ionization chamber for monitoring of the beam during scanning has been constructed. The details of the dose monitors and the position sensitive multi-wire ionization chamber are presented in this work. Furthermore, we have established procedures for verification measurements to ensure the quality of the beam and also methods for calibration of the beam monitors and relative and absolute dosimetry for complex scanned beams. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|