Cleveland Clinic PediPump Lamb Cadaver Fitting Studies.

Autor: Cingoz, Faruk, Fukamachi, Kiyotaka, Ootaki, Yoshio, Kamohara, Keiji, Akiyama, Masatoshi, Ootaki, Chiyo, Kopcak, Michael W., Liu, Jenny, Noecker, Angela, Dudzinski, Dave, Smith, William A., Duncan, Brian W.
Předmět:
Zdroj: Artificial Organs; May2007, Vol. 31 Issue 5, p405-408, 4p, 2 Diagrams, 2 Charts
Abstrakt: We are developing the PediPump, a magnetically suspended, mixed-flow pump, as an implantable pediatric ventricular assist device (VAD). Lamb cadaver fitting studies were performed to determine the optimal pump location and optimal design of the inflow and outflow conduits for chronic in vivo studies. A prototype of the PediPump right and left ventricular assist devices (RVAD and LVAD, respectively) were implanted via a sternotomy or left thoracotomy in four lamb cadavers (3.7–34.2 kg). Via a sternotomy, the RVAD and LVAD required long inflow cannulas when placed into the right or left thoracic cavities, respectively. Via a left thoracotomy, with both pumps implanted in the left thoracic cavity, the RVAD required a 70° inflow cannula and a bent outflow graft while the LVAD required a 130° inflow cannula and a straight outflow graft. In conclusion, left thoracotomy provided optimal fitting for both LVAD and RVAD for the small lamb model. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index