Purpose: The purpose of this study was to evaluate the diagnostic accuracy (sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV)) of the PCR-based BioFire® Joint Infection Panel (BJI Panel) against microbiological culture growth for patients suspected of having a native or prosthetic joint infection. Methods: Synovial fluid and tissue biopsies were prospectively collected from patients from June 2022 to June 2023. The results of the BJI Panel were compared with those of culture growth. Results: 51 samples were included. Including all pathogens, the sensitivity was 69%, the specificity 89%, the PPV 73% and the NPV 86%. Including only pathogens in the BJI Panel, the sensitivity was 100%, the specificity 90%, the PPV 73% and the NPV 100%. Conclusion: The BJI Panel has a high accuracy for detecting the pathogens in its panel, but the absence of important common pathogens from the panel reduces its sensitivity and NPV. With a short turnaround time and precise pathogen detection, the BJI Panel has the potential to add value as a complementary diagnostic method. [ABSTRACT FROM AUTHOR] |