Autor: |
Zanoni, Andrea, Mouloudakis, Kostas, Tayler, Michael C. D., Corrielli, Giacomo, Osellame, Roberto, Mitchell, Morgan W., Lucivero, Vito Giovanni |
Předmět: |
|
Zdroj: |
Journal of Applied Physics; 10/14/2024, Vol. 136 Issue 14, p1-6, 6p |
Abstrakt: |
We demonstrate a sensitive optically pumped magnetometer using rubidium vapor and 0.75 amg of nitrogen buffer gas in a sub-mm-width sensing channel excavated by femtosecond laser writing followed by chemical etching. The channel is buried less than 1 mm below the surface of its fused silica host material, which also includes reservoir chambers and micro-strainer connections, to preserve a clean optical environment. Using a zero-field-resonance magnetometry strategy and a sensing volume of 2.25 mm 3 , we demonstrate a sensitivity of ≈ 1 pT / Hz at 10 Hz. The device can be integrated with photonic structures and microfluidic channels with 3D versatility. Its sensitivity, bandwidth, and stand-off distance will enable detection of localized fields from magnetic nanoparticles and μ L NMR samples. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|