Performance of a Norfentanyl Immunoassay in Specimens with Low Concentrations of Fentanyl and/or Norfentanyl.

Autor: Uljon, Sacha, Tolan, Nicole V., Mahowald, Grace K., Khaliq, Tahira, Urwiller, Elizabeth D., Daluz Fernandes, Maria, Basu, Sankha S., Kang, Phillip, Erickson, Timothy B., Hayes, Bryan D., Chai, Peter R., Melanson, Stacy E. F.
Předmět:
Zdroj: Journal of Applied Laboratory Medicine; Sep2024, Vol. 9 Issue 5, p895-904, 10p
Abstrakt: Background: Many fentanyl immunoassays are limited in their ability to detect norfentanyl. Urine specimens collected from individuals who have been exposed to fentanyl frequently have detectable concentrations of norfentanyl (≥2 ng/mL) but low concentrations of fentanyl (<2 ng/mL) by LC-MS/MS. The Lin-Zhi Fentanyl II Immunoassay (Lin-Zhi) claims 100% cross-reactivity with norfentanyl and therefore may detect exposure missed by other assays. Methods: In addition to verifying the manufacturer's analytical sensitivity claims, we selected 92 urine specimens with low-positive Lin-Zhi results (1-99 absorbance units, lowest 10%) for analysis by the Immunalysis Health Equity Impact Assessment and ARK II fentanyl methods. The accuracy of the 3 immunoassays was compared to LC-MS/MS as the reference method. Results: Spiking studies using purified fentanyl and norfentanyl and a set of 100 consecutive specimens confirmed the manufacturer's claims of limit of detection for fentanyl (3.8 ng/mL) and norfentanyl (5.0 ng/mL). However, the 92 low-positive patient specimens demonstrated concentrations of norfentanyl and fentanyl below 2.0 ng/mL by LC-MS/MS, with 47 (51%) having only norfentanyl detected. When comparing Lin-Zhi to the Immunalysis and ARK II immunoassays, only 27 (29%) of the 92 specimens were concordant. Fifty-two (57%) of the specimens were positive by LC-MS/MS and Lin-Zhi but false negative by one or both other immunoassays. Seven specimens (8%) were positive by Lin-Zhi but negative by the other immunoassays and had undetectable concentrations (<2 ng/mL) of fentanyl and norfentanyl by LC-MS/MS. Conclusions: The clinical sensitivity of the Lin-Zhi exceeds the manufacturer's claims, providing results comparable to LC-MS/MS methods. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index