Autor: |
Dwivedi, Pooja, Jha, Ashish Kumar, Mithun, Sneha, Sawant, Viraj, Vajarkar, Vishal, Chauhan, Manoj, Choudhury, Sayak, Rangarajan, Venkatesh |
Předmět: |
|
Zdroj: |
Radiation Protection Dosimetry; Sep2024, Vol. 200 Issue 14, p1384-1390, 7p |
Abstrakt: |
This study aimed to evaluate the dose in different protocols of 18F-2-fluoro-2-deoxy-D-glucose positron emission tomography/computed tomography (PET/CT) procedure. The retrospective study involves 207 patients with confirmed malignancies who underwent PET/CT. Effective dose (E) from PET was estimated based on injected activity and dose coefficient as per International Commission on Radiation Protection (ICRP) 128. Estimation of E from CT was done utilizing the dose length product (DLP) method and conversion factors as per ICRP 102. There was a significant statistical difference observed in E between different PET/CT protocols (P <.001). E of PET in the whole body (WB) was found to be 4.9 ± 0.9 mSv, whereas mean volume computed tomography dose indexvol, DLP, and E of CT in WB were 7.0 ± 0.2 mGy, 674.3 ± 80.7 mGy.cm, and 10.1 ± 1.2 mSv, respectively. No linear correlation was seen between the size-specific dose estimate and E of CT (r = −0.003; P =.978). The total mean E in WB PET/CT was 17.0 ± 1.7 mSv. CT dose was contributing more than PET dose in all protocols except brain PET/CT. Optimization strategies can be evaluated only if monitored periodically. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|