Autor: |
Kim, Hwajung, Kim, Soohyun, Park, Soyoon, Kim, Sunhwa, Choi, Young, Kim, Ju Youn, Oh, Yong-Seog, Kim, Sung-Hwan |
Zdroj: |
Journal of Interventional Cardiac Electrophysiology; Aug2024, Vol. 67 Issue 5, p1211-1217, 7p |
Abstrakt: |
Background: The avascular capsule around the generator of the cardiac implantable electronic device (CIED) could be susceptible to bacterial colonization and source of infection. Capsulectomy during CIED generator replacement may be beneficial in preventing device infection, but there is a lack of evidence. Methods: This prospective randomized trial, conducted from December 2013 to December 2019, included 195 patients divided equally into two groups. In the intervention group (n = 97), capsule removal was performed on the floor of the pocket, while it was not performed in the control group (n = 98). In both groups, swab culture was performed in the pocket. The primary outcome was the occurrence of device infection requiring pocket revision. Results: A total of 195 patients were included (mean age 70.2 ± 13.6 years, 55.4% women), with an average follow-up period of 54.3 ± 28.9 months. Among 182 patients undergoing microbiological cultures of pockets, 19 (10.4%) were confirmed positive, and Staphylococcus species were identified most frequently. The primary outcome occurred in 4 (2.1%), and there was no significant difference between the two groups (3.1% vs. 1.0%, p = 0.606). Hematoma has occurred in 10 patients (3.1% vs. 7.1%, p = 0.338), one of them required wound revision. In multivariable analysis, the occurrence of hematoma was the only independent risk factor associated with device infection (HR 13.6, 95% CI 1.02–181.15, p = 0.048). Conclusions: In this long-term prospective study, capsulectomy during the replacement of the generator did not reduce the incidence of device infection. There was no association between bacterial colonization in the capsule around the generator and CIED infection. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|