Deep vein thrombosis is a condition that can cause life-threatening complications if left untreated. In addition to traditional medical treatment approaches for its treatment, it is included in angiographic methods with a high success rate, which are widely used today and give results in a shorter time. Angiographic methods have serious complications related to the materials used and angiographic techniques. In our case, pharmacomechanical thrombectomy was performed with endovenous angiographic method due to acute iliofemoral deep vein thrombosis. During the removal of the poseidon hydrophilic guide piece, which broke off in the mid-femoral region during the procedure, it moved proximally and was attached to the vena cava filter. The broken guidewire piece was successfully removed together with the vena cava filter. Our case was found worthy to be presented in order to emphasize the importance of using vena cava filter in such interventions. [ABSTRACT FROM AUTHOR] |