Autor: |
Simpson, Andrea, Hersbach, Adam A., McDermott, Hugh J. |
Předmět: |
|
Zdroj: |
International Journal of Audiology; May2005, Vol. 44 Issue 5, p281-292, 12p, 1 Diagram, 3 Charts, 5 Graphs |
Abstrakt: |
The performance of an experimental frequency compression hearing device was evaluated using tests of speech understanding in quiet. The device compressed frequencies above a programmable cut-off, resulting in those parts of the input signal being shifted to lower frequencies. Below the cut-off, signals were amplified without frequency shifting. Subjects were experienced hearing aid users with moderate-to-severe sensorineural hearing loss and sloping audiograms. Their recognition of monosyllabic words was tested using the experimental device in comparison with conventional hearing aids. Of the 17 subjects, eight showed a significant score improvement (p? [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|