Autor: |
Mussmann, Bo, Skov, Peter Marshall, Lorentzen, Morten H, Skjøt-Arkil, Helene, Graumann, Ole, Andersen, Michael B, Jensen, Janni |
Předmět: |
|
Zdroj: |
Acta Radiologica Open; Mar2023, Vol. 12 Issue 3, p1-5, 5p |
Abstrakt: |
Background: In suspected community-acquired pneumonia (CAP), chest CT is superior to the routinely obtained radiographs (CXR), but administers higher radiation doses. However, ultra-low-dose CT (ULDCT) has shown promising results. Purpose: To compare radiation dose and image quality using standard and ULDCT protocols designed for a multicenter study encompassing three CT scanner models from GE, Canon, and Siemens. Material and methods: Patients with suspected CAP were referred for non-contrast standard dose chest CT (NCCT) and ULDCT. Effective radiation dose and Contrast-to-Noise Ratio (CNR) was calculated. Results: Mean effective doses were GE (n = 10) 6.93 mSv in NCCT and 0.27 mSv in ULDCT; Canon (n = 9) 3.48 in mSv NCCT and 1.11 mSv in ULDCT; Siemens (n = 10) 2.85 mSv in NCCT and 0.45 mSv in ULDCT. CNR was reduced by 29–39% in ULDCT. Conclusion: The proposed CT protocols yielded dose reductions of 96%, 68%, and 84% using a GE, Canon, and Siemens scanner, respectively. [ABSTRACT FROM AUTHOR] |