Evaluation of the oxygen transfer performance of a small membrane oxygenator.

Autor: Esipova, O. Yu., Buchnev, A. S., Drobyshev, A. A., Kuleshov, A. P., Grudinin, N. V., Bogdanov, V. K.
Předmět:
Zdroj: Biomedical Engineering; Nov2023, Vol. 57 Issue 4, p260-264, 5p
Abstrakt: A series of small-sized membrane oxygenators (N = 10) has been developed with 90–100 aligned microporous polypropylene hollow fibers placed within a sheath hermetically sealed at both ends to isolate the perfusion solution. A hydrodynamic set-up was assembled to evaluate oxygenator performance in terms of gas and perfusate flow, perfusion pressure, and temperature drop over the range of flow rates 5–70 ml/min. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index