Autor: |
Kresh, Yasha M., Parsonnet, Victor, Myers, George H., Parsonnet, Brian |
Předmět: |
|
Zdroj: |
Pacing & Clinical Electrophysiology; May1979, Vol. 2 Issue 3, p297-304, 8p |
Abstrakt: |
A method of measuring and analyzing beat-to-beat interval stability of implanted pacemakers has been devised, and a number of normal and failing pacers have been studied in this way. It was seen that normal pacer rates were remarkably stable, but that abnormalities in beat-to-beat interval were found in 67.4% of all failing pacers. The abnormalities consisted primarily of excessive interval (rate) scatter, linear and cyclic fluctuation, and transients following application of a magnet. It is expected that in problem cases a heat-to-beat analysis will be of help in differentiating normal from abnormal pacer function, and will provide confirmation signs of pacer failure: it may also uncover unusual and heretofore unpredictable failure modes. [ABSTRACT FROM AUTHOR] |