Autor: |
Paraskevaidis, Stelios, Mochlas, Sotirios, Hadjimiltiadis, Stavros, Louridas, Georgios |
Předmět: |
|
Zdroj: |
Pacing & Clinical Electrophysiology; Apr1999, Vol. 22 Issue 4, p689-690, 2p, 1 Diagram |
Abstrakt: |
We present the case of a patient with a DDD pacemaker and intermittent P wave sensing due to T wave oversensing by the ventricular lead. The T wave sensing caused initiation of an extended atrial refractory period and the P waves, falling within this period, were not sensed. The problem was solved by decreasing the ventricular sensitivity. [ABSTRACT FROM AUTHOR] |