Autor: |
Burdonov, K., Lelièvre, R., Forestier-Colleoni, P., Ceccotti, T., Cuciuc, M., Lancia, L., Yao, W., Fuchs, J. |
Předmět: |
|
Zdroj: |
Review of Scientific Instruments; Aug2023, Vol. 94 Issue 8, p1-6, 6p |
Abstrakt: |
We present the design and absolute calibration of a charged particle online readout CMOS system tailored for high-power laser experiments. This system equips a Thomson parabola spectrometer, which is used at the Apollon petawatt scale laser facility to measure the spectra of protons produced by high-intensity laser–target interactions. The RadEye1 CMOS matrices array detectors are paired with a custom triggering system for image grabbing. This allows us to register the proton and ion signals remotely. The repetition rate is presently 1 shot/min, but the frame grabbing enables the system to be compatible with modern high-power lasers running, e.g., at 1 Hz. We detail here the implementation, in the harsh electromagnetic environment of such interactions, of the system, and its absolute calibration, which was performed for proton energies from 4 to 20 MeV. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|