Inadvertent Transarterial Pacemaker Insertion: An Unusual Complication.

Autor: Lepore, Vicenzo, Pizzarelli, Giampaolo, Dernevik, Leif
Předmět:
Zdroj: Pacing & Clinical Electrophysiology; Jul1987, Vol. 10 Issue 4, p951-954, 4p
Abstrakt: We describe an unusual complication of pacemaker treatment in a patient who died after a replacement operation. In a difficult situation in which a functioning pacemaker was highly desirable and in which most of the available veins had already been used, the pacemaker electrode was inserted, by mistake, through a small artery. This was not detected by fluoroscopy during surgery. The postoperative X-ray examination seemed to indicate that the electrode tip was located in the coronary sinus, but the subsequent autopsy revealed it to be located in the left ventricle. [ABSTRACT FROM AUTHOR]