Autor: |
Huang, Qiuchen, Cai, Zhiqiang, Li, Wei, Chen, Ruonan, Zhang, Wei, Jin, Keyun, Zhao, Yi, Li, Yiwen, Sun, Tao |
Předmět: |
|
Zdroj: |
Analytical Letters; 2023, Vol. 56 Issue 11, p1713-1726, 14p, 1 Diagram, 3 Charts, 6 Graphs |
Abstrakt: |
A novel calix[6] arene polymer stationary phase (C6A-PEG) is reported that combines the π-electron-rich cavity of calix[6] arene (C6A) with polyethylene glycol (PEG) units for gas chromatography (GC). Thermogravimetric analysis revealed a 5% weight loss of C6A-PEG at 318 °C, and its static-coated capillary column showed high efficiency of 2924 plates/m (k = 3.48) determined by 1-octanol at 120 °C. As a result, the moderate polar C6A-PEG column showed clearly advantageous performance for the Grob reagent and a mixture of 25 analytes with high selectivity for hydrogen-bonded acids. Also, it exhibited high resolving performance for various isomer mixtures, in particular high-resolution performance (R > 2.2) for the challenging toluidine and dimethylaniline systems with distinct advantages compared to the C6A and PEG columns (R < 1.5 for some pairs). This work demonstrates the potential of calixarene-based polymers as novel stationary phases for gas chromatographic analysis. [ABSTRACT FROM AUTHOR] |