Angiojet™ mechanical thrombectomy‐induced hemolysis triggering blood‐leak alarms.

Autor: Selvananthan, Babitha, Aziz, Omar, Lee, Arvind D., Werner‐Gibbings, Keagan, Bose, Bhadran, Sud, Kamal
Předmět:
Zdroj: Hemodialysis International; Apr2023, Vol. 27 Issue 2, pE19-E22, 4p
Abstrakt: Blood leak alarms are increasingly rare and may be triggered by rupture of the dialyzer membrane, or intravascular hemolysis. We report two patients who developed hemolysis following Angiojet™ thrombolysis and thrombectomy of thrombosed arteriovenous fistulas, triggering blood leak alarms on hemodialysis, the occurrence of which has not been reported before. AngioJet™‐induced hemolysis should be considered in the differential diagnosis for blood leak alarms occurring soon after an intervention on an arterio‐venous (AV) fistula. [ABSTRACT FROM AUTHOR]